SP Genevac ATS
HT Series 3i

SP Genevac HT Series 3i Vacuum Evaporator

Brochures
More Info

Specifications

Evaporator
Max. Rotor Speed
1415 rpm
Max. load per swing
1.5 kg
Max. operational imbalance
80 g
Dimensions (W x D x H) 1
660 x 710 x 840 mm
Weight (approx.)1
193.3 kg
Vacuum Pump (Integral)
Type
Oil-free Scroll
Ultimate system vacuum
< 4mbar
Dimensions (WxDxH)
432 x 282 x 302 mm
Weight
26.2 kg
Vacuum hose/control cable
3 m
Max. coolant inlet pressure
30 psi (2 bar g)
Condenser
Type
Dual-stage vapour compression
Refrigerant gas - stage 1
R449A
Refrigerant charge - stage 1
320 g
Refrigerant GWP - stage 1
1,397
Refrigerant CO2e - stage 1
< 0.001 tonnes
Total CO2 equivalent (CO2e)
0.5 tonnes
Ultimate low temperature2
-75 °C
Max. Pressure (PS)
30 bar
Emissions
Noise (@ 1 metre)
65 dB(A)
Exhaust hose (supplied)
6 mm ID/8 mm OD
Electrical
Supply
230 V 50 Hz
220 V 60 Hz
208 V 60 Hz
Max. supply input
1500 A
Power Consumption
Current (A) at unit voltage
Operational Environment
Ambient temperature
15 °C to 30 °C
Relative humidity
10-80% non-condensing
Altitude
Sea-level to 1600 m
Min. ventilation air-gap
70 mm
Installation environment
Indoor only.
Static-dissipative laboratory or similar
Solvent Capacity & ACC Range
Max. solvent capacity
4.5 L
Refridgeration ACC range
100 °C
Inert Gas Supply Requirements
Max. pressure
2 bar g (3 bar abs.)
Min. pressure
1.5 bar g (2.5 bar abs.)
Flow rate (nominal)
50 litres/min @ STP
Hose length
2.5 m
Max. consumption (purge)
120 litres approx.
Max. consumption (blanket)
60 litres/hour approx.
Connector type
3/8" BSP female
New nXDS6i Vacuum Pump

Provided with method specified pump purging to optimize vacuum and pump reliability

1 Varies with build options
2 Ultimate low temperature; operational values may vary
3 3-10 °C permissable during transport (only)